Development of the SRK/T intraocular lens implant power calculation formula

您所在的位置:网站首页 srk skx Development of the SRK/T intraocular lens implant power calculation formula

Development of the SRK/T intraocular lens implant power calculation formula

#Development of the SRK/T intraocular lens implant power calculation formula| 来源: 网络整理| 查看: 265

A new implant power calculation formula (SRK/T) was developed using the nonlinear terms of the theoretical formulas as its foundation but empirical regression methodology for optimization. Postoperative anterior chamber depth prediction, retinal thickness axial length correction, and corneal refractive index were systematically and interactively optimized using an iterative process on five data sets consisting of 1,677 posterior chamber lens cases. The new SRK/T formula performed slightly better than the Holladay, SRK II, Binkhorst, and Hoffer formulas, which was the expected result as any formula performs superiorly with the data from which it was derived. Comparative accuracy of this formula upon independent data sets is addressed in a follow-up report. The formula derived provides a primarily theoretical approach under the SRK umbrella of formulas and has the added advantage of being calculable using either SRK A-constants that have been empirically derived over the last nine years or using anterior chamber depth estimates.